7.6
Required Aids
Reprocessing step Items Needed
Precleaning
Soft plastic brush/soft cloth
10 ml disposable syringe
Irrigation tube
Machine cleaning/
Irrigation tube
disinfection
7.7
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times). You can mount an irriga-
tion tube to the irrigation connection of the electrode shaft
for this purpose.
7.8
Cleaning and disinfection by machine
The washer/disinfector must have fundamentally certified ef-
fectiveness (e.g. CE marking according to DIN EN ISO 15883).
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11. Prohibited ingredients: organic solvents, oxidizing
agents, halogens, aromatic/halogenated hydrocarbons.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
26