Section 2- Operation

Installation and Verification
● Battery installation
Caution: The Pulse Oximeter does not operate with dead batteries and
can not be powered by external power source. Install new batteries.
1. Unplug all accessories from the Pulse Oximeter, and press the menu button to
access the Setting Interface, turn the PULSE OXIMETER off. See table 1.
2. Remove the battery rear cover towards the direction indicated on the cover.
3. Insert two "AAA" size batteries, making sure the battery's positive and negative
poles are correctly oriented in the battery compartment as shown in Figure 5.
4. Close the battery rear cover.
Battery rear cover
Figure 5: Installing Batteries
● Performance Verification
1. Performance Tests
Power up the device to get the oximeter ready for measurement.
2. Power-On Self-Test
Before using the PULSE OXIMETER, you must verify that the PULSE OXIMETER is
working properly and is safe to use. Proper working conditions are verified each time
when the PULSE OXIMETER is turned on as described in the following procedure. The
verification procedure (POST) takes 2 to 3 seconds to complete.
Caution: If any indicator or display element does not light when the PULSE OXIMETER
is turned on, do not use the PULSE OXIMETER. Instead, contact qualified service
personnel, your local MED-LINKET representative, or MED-LINKET's Technical Services
Department.
Note: The Pulse Oximeter automatically starts the Power-On Self-Test (POST) to
ensure that its internal circuits are functioning properly.