No display on
screen, but the
transmitting tube of
LED lights on.
No reading display
on Pulse-Oximeter
Fail auto-off
Inaccurate Temp
Measurement

Section 4- Electromagnetic Environment

Electromagnetic Interference Caution

This Fingertip pulse oximeter is not designed for use in environments in which the
pulse can be obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the device may not seem to operate
correctly.

Electromagnetic Environment

The PULSE OXIMETER is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or user of the PULSE
OXIMETER can help prevent electromagnetic interference by maintaining a mini-
mum distance between portable and mobile RF communications equipment (trans-
mitters) and the PULSE OXIMETER as recommended below, according to the maxi-
mum output of the communications equipment.
Warning:
PULSE OXIMETER should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, PULSE OXIMETER should be
observed to verify normal operation in the configuration in which it will be used.
With damage in display screen or
break in the connection spot of
display screen
Poor perfusion problem (general-
ly, there's no display on pulse
amplitude indicator but the light
of LED transmitting tube is on
and the finger is inserted in
place)
The transmitting tube of LED
lights off
External Temp-probe is
still working
Damage in collection tube or
other device parts.
The surface Temp-Probe did not
firmly stick on the skin surface.
Wrong position of Temp-Probe
No enough measuring time
English 18
Contact authorized
distributors
If there's no pulse amplitude
displayed on the screen,
Please adjust the finger
position;Use your middle or
index finger in preference;
Warm your fingers;
Contact authorized
distributors
Pull out the external
temp-probe
Contact authorized
distributors
Stick the Temp-Probe on the
proper measuring position
by medical proof fabric
Place the Temp-Probe in
proper position according
to the specification
Keep the correct measuring
method by 10 mins, and then
get the result.