SpO
Short press the button in the interface
of "Alm Setup 2", move the cursor to
SpO
Low
select the item of "SpO2 Alm Lo", long
2
Limit setting
press the button to adjust the
parameter of SpO2 in the scope of 50%
to 98%.
The default lower limit is 94%.
Short press the button in the interface
of "Alm Setup 2", move the cursor to
Pulse Rate (PR)
select the item of "PR Alm Hi", long
High Limit
press the button to adjust the parame-
setting
ter of BPM in the scope of 32-245bpm.
The default upper limit is 130 bpm.
Short press the button in the interface
of "Alm Setup 2", move the cursor to
select the item of "PR Alm Lo", long
Pulse Rate (PR)
Low Limit
press the button to adjust the
setting
parameter of BPM in the scope of
30-243bpm. The default lower limit is
50bpm.
Short press the button in the interface of
"Sounds Setup", move the cursor to select
Temp High
the item of "Temp Alm Hi", long press the
Limit setting
button to adjust the parameter of tem-
perature in the scope of 78-113
Short press the button in the interface of
"Sounds Setup", move the cursor to select
Temp low
the item of "Temp Alm Lo", long press the
Limit setting
button to adjust the parameter of tem-
perature in the scope of 77-112
After completed above setting, press the
button switch to any interface of setting,
Return to the
POST display
move the cursor to select the item of
"Exit" to return to the POST display.
Low Limit setting
2
English 16
.
℉
.
℉
97
75
%SpO
BPM
2