Problem
Possible cause
The intensity on
the device is set
too high.
Water contact
electrodes are not
Stimulation is
sufficiently moist.
uncomfort-
able.
Connection cable
is worn/faulty.
Water contact
electrodes are
worn/faulty.
Irregular
Connection cable
stimulation.
faulty.
Water contact
Stimulation is
electrodes are not
ineffective.
positioned cor-
rectly on the skin.
Solution
Lower the intensity
on the device.
Switch off the de-
vice. Moisten elec-
trodes.
Replace the con-
nection cable.
Replace the univer-
sal cuff.
Reduce the inten-
sity and rotate the
connection cable on
the socket by 90°.
If additional power
failures occur, re-
place the connec-
tion cable.
Remove water
contact electrodes
and reposition on
the skin. Ensure
there is sufficient
contact with the
skin.
29
Problem
Possible cause
The universal cuff
is not positioned
correctly on the
skin.
The universal cuff
The skin
is dirty.
turns red
and/or a
stabbing pain
One of the
occurs.
water contact
electrodes has a
scratch.
Water contact
electrodes are not
sufficiently moist.
The universal cuff
has worked itself
loose from the
skin.
Stimulation
The connection
stops during
cable has worked
use.
itself loose.
The batteries are
empty.
Solution
Ensure that the
universal cuff sits
securely on the skin
and cannot move.
Clean the universal
cuff as described in
these instructions
for use.
Replace the univer-
sal cuff.
Switch off the de-
vice. Moisten elec-
trodes.
Stop the application
and reposition the
universal cuff.
Stop the application
and reinsert the
connection cable.
Replace the batter-
ies on the control
unit.