DeVilbiss CPAP Nasal 9000 Guía De Instrucciones página 5

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

8
During a single fault, the blower is maintained within
spec or at less than the maximum limited pressure. For
a firmware detected fault, the blower is set to a fail
safe mode in which the blower is turned off and the
detected error code is shown on the LCD display. CO
rebreathing is minimized by keeping the mask vent
holes open at all times.
1. Plug CPAP into AC power using appropriate power
cord. Will work on 100-240 volts AC (50 or 60
Hz) with no changes except for the end of the
power cord where it plugs into the wall.
2. The display will show Set Pres and the CPAP will
increase to the prescribed CPAP pressure.
3. The display shows the CPAP pressure prescribed
by the doctor.
NOTE—Your physician has chosen the appropriate
delay time (0, 10, 20, 30 or 45 minutes) that
has been preset by your Sunrise Medical
provider or clinician.
4. Press the Delay button (Figure 8) to start the
pressure delay cycle if prescribed. Pressure will
decrease to the Delay Pressure setting and return
to the prescribed Set Pressure after the Delay
Time has expired.
5. When you awake in the morning, press the ON/OFF
button to turn the blower OFF (Figure 9).
Model 9000 Altitude Selection
Model 9000 has manual altitude selection. This allows
the patient to correct the CPAP pressure when traveling.
To correct the CPAP pressure for altitude changes:
1. Plug CPAP into AC power using appropriate power
cord. Will work on 100 to 240 volts AC (50 or 60
Hz) with no changes except for the end of the
power cord where it plugs into the wall.
2. The display will show Set Pres and the CPAP will
increase to the prescribed CPAP pressure.
3. Press the Select button on the keypad until the
LCD display shows Altitude.
A-900
All manuals and user guides at all-guides.com
O p e r a t i o n
LCD Display
Set Pres
8.0 cmH
2
Delay button
Figure 8
ON/OFF
Select button
Figure 9
Altitude
LCD Symbol
1 ❙
0-2500
2 ❙ ❙
2500-5000
3 ❙ ❙ ❙
5000-7500
4 ❙ ❙ ❙ ❙
7500-9000
O p e r a t i o n
4. Use the UP or DOWN keys on the keypad to change the altitude setting.
5. Operate the CPAP normally.
6. Remember to return Altitude to its original setting upon returning home.
Model 9000 LCD Display Window Definitions
Pressing the SELECT button changes the display to show information about the CPAP.
Press the UP or DOWN button to change the displayed setting. (The patient can only
change the Altitude and Language.)
• Set Pressure– CPAP pressure (cmH
• Altitude Range– Patient selectable, corrects CPAP pressure for changes in altitude
while traveling. See Operating Instructions for ranges and use.
• Delay Time– Time for the CPAP to reach the prescription pressure after pressing
the Delay button.
• Delay Pressure– Pressure delivered during the delay cycle.
• Compliance Hours– Number of hours recorded while breathing was present.
• Unit Hours– Number of hours recorded while CPAP blower was ON.
• Firmware Version– Identification for firmware installed in the CPAP.
• Serial Number– Serial number of the CPAP unit, same as on the label.
• Language– Language selected for LCD display messages.
• Off- Blower is OFF AC or DC power is connected.
• Error- An error has been detected by the firmware, the CPAP is in a fail safe mode.
OPERATION OF THE 9001
CO
Rebreathing
2
CO
rebreathing is minimized during use by providing 3cmH
2
which creates airflow through mask vent holes. It is important that mask vent holes
are not blocked.
During a single fault, the blower is maintained within spec or at less than the
maximum limited pressure. For a firmware detected fault, the blower is set to a fail
safe mode in which the blower is turned off and the detected error code is shown on
the LCD display. CO
rebreathing is minimized by keeping the mask vent holes open at
2
all times.
1. Plug CPAP into AC power using appropriate power cord. Will work on 100 to 240
volts AC (50 to 60 Hz) with no changes except for the end of the power cord
where it plugs in the wall. Model 9001 will also operate on 12 VDC.
2. The display will show Set Pres and the CPAP will increase to the prescribed CPAP
pressure.
3. The display shows the CPAP pressure prescribed by the doctor.
4. Press the Delay button to start the pressure delay cycle, if prescribed. Pressure will
decrease to the Delay Pressure setting and return to the prescribed Set pressure
after the Delay Time has expired (Figure 8).
5. When you awake in the morning, press the ON/OFF button to turn the blower OFF
(Figure 9).
0) prescribed by your doctor
2
O minimum pressure,
2
9
A-900

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cpap nasal 9001

Tabla de contenido