Calibration Standards; T-Cal Standards - Correct Storage And Using - Tintometer Lovibond TB 300 IR Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 103
<NTU adjustment>
store

2.3.6 Calibration standards

The T-CAL standards purchased with the instrument should be used to calibrate the instrument.
It is possible to perform calibration with fresh dilutions of a Formazin Stock Solution. Using
dilutions of a Formazin Stock Solution it is recommended to calibrate and measure with an
indexed single vial or a set of matched vials.
2.3.6.1 T-CAL standards – correct storage and using
For optimum results when using T-CAL standards, adhere to the following recommenda-
tions:
Storing T-CAL standards
• T-CAL standards should always be stored in their original vial.
• Preferably store the vials upright.
• Store standards between 5 and 25°C.
• Avoid prolonged exposure to temperatures exceeding 35°C.
• Store away from direct sunlight.
• Always allow the standards to acclimatise to ambient instrument temperature before use
(not to exceed 35°C).
• T-CAL standards have a shelf life of 12 months if stored correctly.
Using < 0.1 NTU Standards
Caution: Never shake or invert this < 0.1 NTU standard.
• After delivery allow the vial to stand for at least 24 hours before use.
• If the standard has been shaken it can take several hours for all the bubbles to dissipate
and we suggest the vial is left for at least 24 hours as outlined above.
• If the standard has been inverted accidentally wait at least 15 minutes before using.
Preparing and using standards – infrequent use
Note: These instructions apply to all standards except the < 0.1 NTU standard. If you are
using the < 0.1 NTU standard follow the instructions given above.These instructions apply to
standards that have been sitting undisturbed for more than one week or to new standards.
1. Shake the standard vigorously for 2-3 minutes.
2. Let the vial stand undisturbed for 5 minutes.
3. Invert the vial between 5-10 times.
4. Immediately place the vial in the sample chamber and wait for 1 minute (countdown).
Preparing and using standards – frequent use
Note: These instructions apply to all standards except the < 0.1 NTU standard. If you are
using the < 0.1 NTU standard follow the instructions given above.These instructions apply to
frequently used standards (daily or weekly usage).
44
Press [ ] key to store.
Pess [Esc] key to cancel the calibration.
TB300 IR 6 03/2022

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

194000-b194000

Tabla de contenido