Fault Indication - GERATHERM Duotemp GEF-10 Instrucciones De Servicio

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 68

FAULT INDICATION

FAULT OR
FAULT
SYMBOL
No display on
GB
the LCD panel
Measurement
not possible (or
an abnormal
value is dis-
played)
An abnormal
temperature
value is dis-
played.
LO or HI sym-
bol is displayed
12 - GB
FAULT DE-
SCRIPTION
The battery has
run out.
Incorrect bat-
tery polarity.
The thermom-
eter is not
ready.
The probe tip
is dirty or dam-
aged. Did you
hear the beep
after pressing
the ON button?
The tempera-
ture measured
is outside the
measuring
range
(0 °C -100 °C
or
32 °F-212°F)
Low:
temperature
too low
High:
temperature
too high
CORRECTIVE
MEASURE
Replace the
battery. Please
note:
The (+) side
of the battery
must face
upwards.
Wait until the
°C symbol is
displayed.
Clean the probe
tip or get it
repaired. Wait
until you hear
the beep before
removing the
thermometer
from the ear or
forehead.
Check that the
probe tip is
clean and that
the thermom-
eter is properly
inserted into
the ear or
placed on the
forehead.