7. CLEANING AND DISINFECTION

WARNING: Ensure that the illuminator is de-energized and disconnected from mains power before
attempting to clean and disinfect.
The illuminator can be wiped down with commercially available cleansers commonly used for
disinfection of electronic equipment in hospitals such as ethyl or isopropyl alcohols, disinfecting sprays
containing quaternary ammonium compounds, or hydrogen peroxide.
WARNING: DO NOT use strongly caustic or acidic cleansers such as "Clorox" hypochlorite bleach,
ammonia, muriatic acid, or similar products. DO NOT use acetone, methyl ethyl ketone, or halogenated
/ chlorinated hydrocarbon solvents or cleansers containing any of these restricted compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids onto the device. Do not
allow liquids to enter the device seams or ventilation openings.
Follow all applicable bloodborne pathogen procedures as required by OSHA and/or your hospital
when cleaning and disinfecting the product.
WARNING: The illuminator is not sterilizable. DO NOT attempt to autoclave the device.
8. MAINTENANCE, SERVICING, REPAIR & WARRANTY
Performance of preventative maintenance is not essential. Defective items or equipment are to be
serviced and repaired exclusively by persons authorized by the manufacturer. All repair work shall
employ original manufacturer's parts only.

8.1 Lamp Replacement

WARNING: Ensure that the illuminator is unplugged from mains power and de-energized before
attempting to replace the lamp.
WARNING: Ceramic Xenon lamps are at high internal pressure when cold and at operating
temperature. Ceramic Xenon lamps may unexpectedly rupture resulting in discharge of hot fragments
of quartz and/or glass and metal. Only handle lamps with protective covers in place.
WARNING: Do not change lamp while touching patient.
NOTE: This illuminator is designed for exclusive use with the Sunoptic Surgical P/N: SSX0400 lamp
module. Failure to use this lamp could void warranty.
Turn unit off and unplug from mains power. Open lamp door located on side of illuminator.
Move the lamp positioning lever (found to the right of the lamp) clockwise, from horizontal to vertical,
and pull the lamp cartridge out.
Grasp only the horizontal top and bottom grip flanges of the plastic housing of the lamp, pull straight
out to remove the lamp. Insert a new one (lamp pointed to the left), pushing firmly to assure full
engagement to the power connectors. Lower the lamp engagement lever counterclockwise, from
vertical to horizontal, and observe that the lamp rises about 3mm as it moves to its proper operating
position.
Log the serial number and hours on the hour meter. Close the lamp compartment door. Re-connect
the power cord and turn the light source on according to 6.1. Reset hour meter to zero by
depressing hour meter reset button on rear panel.
LIT-259 Sunoptic Surgical
Rev. B
®