18
Specifications
Physical Characteristics
DIMENSIONS
WEIGHT
RELATIVE NOISE LEVEL
HOSE
FILTRATION SYSTEM
RECOMMENDED FILTER CHANGE
MOUNTING
Temperature Characteristics
RECOMMENDED OPERATING ENVIRONMENT
TEMPERATURE CONTROL
HEAT GENERATED
OPERATING TEMPERATURES
Safety System
THERMOSTAT
ALARM SYSTEM
OVERCURRENT PROTECTION
Bair Hugger Model 750 Temperature Management Unit Operator's Manual
12˝ high x 13.5˝ deep x 10˝ wide
30 cm high x 34 cm deep x 25 cm wide
15.5 lb (7kg)
55 dBA
Flexible, compatible with the Bair Hugger
brand 241 blood/fluid warming system.
0.2 µm air filter
Every 500 hours of use.
Can be clamped to an IV pole, placed on a
hard surface, or mounted to the rolling stand
accessory.
15°C-25°C
Electronically controlled.
1644 BTU/hr (average), 482 W (average)
Average temperatures at the end of the hose:
HIGH: 43° ± 1.5°C
MED: 38° ± 1.5°C
LOW:
Independent electronic circuit; thermal cut-
off shuts the heater OFF at preset high
temperature of 53°C ± 3°C at the end of the
hose, back-up over-temperature detection at
hose inlet.
Over-temperature (53° ± 3°C ): red
Over-temp indicator light flashes, alarm
sounds, heater and blower shut down,
operating indicator lights turn OFF, control
panel becomes unresponsive.
Fault: amber Fault indicator light flashes,
alarm sounds.
Dual input fused lines.
109.4° ± 2.7°F
100.4° ± 2.7°F
32° ± 1.5°C
89.6° ± 2.7°F