STARTING THE INFUSION
1.
Connect tubing to patient's catheter. Make sure the connection is secure.
2.
Select the appropriate flow rate by turning the dial on the Select-A-Flow until the dial clicks into place,
and the flow rate setting is aligned with the ml/hr mark on the face of the Select-A-Flow. Open clamp.
CAUTION: Dial must "click" into place to ensure accurate flow rate. Flow rate is unpredictable
if it is dialed between rate settings.
CHANGING THE FLOW RATE DURING AN INFUSION
1.
Insert the rate-changing key into the dial.
2.
Turn the dial until the new flow rate is selected. Make sure you hear the dial "click" into place and
the selected flow rate is aligned below the ml/hr mark on the Select-A-Flow.
3.
Remove the key from the dial and put in a safe place for later use.
CAUTIONS:
1.
Do not underfill pump. Underfilling the pump may significantly increase the flow rate.
- Filling the pump less than nominal results in faster flow rate.
- Filling the pump greater than nominal results in slower flow rate.
- Temperature will affect solution viscosity, resulting in faster or slower flow rate. The Easypump C-bloc RA
Select-A-Flow has been calibrated using Normal Saline (NS) as the diluent and room temperature
(22°C, 72°F) as the operating environment. The Select-A-Flow should be worn outside clothing
and kept at room temperature.
Table 1
Nominal Fill Volume (ml)
Maximum FIll Volume (ml)
Retained Volume (ml)
Refrigerator to Room Temp (hr)
Delivery Accuracy: When filled to nominal volume, flow accuracy is ± 20% of the labeled flow rate
when infusion is started 0-8 hours after fill and delivering normal saline as the diluent at 22°C, 72°F.
NOTE: For models including PCA delivery, see PCA product insert for bolus information.
Rx only
= CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
The product is covered by one or more of the following patents: 7,455,072. Additional U.S. and Foreign
Patents may be issued and/or pending.
®
Select-A-Flow and C-bloc are registered trademarks of I-Flow Corporation registered with the U.S. Patents
and Trademark Office.
Easypump is a trademark of B. Braun.
®
270
400
335
550
≤8
≤15
10
12
6
Select-A-Flow
600
models are color-coded.
750
GREEN:
1, 2, 3, 4, 5, 6, 7 ml/hr
≤22
WHITE:
15
2, 4, 6, 8, 10, 12, 14 ml/hr