Molybdenum, With Tablet (Mo 3) - Orbeco HELLIGE SC450 Manual De Instrucciones

Ocultar thumbs Ver también para HELLIGE SC450:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Methods
Molybdate with Tablet
Mo 3
Tablet
Powder
0.6 – 30 mg/l Mo
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
0.0.0
perform zero calibration (see "Operation").
Tablet
Powder
Remove the vial from the sample chamber and empty the vial.
Fill 20 ml of the water sample in a 100 ml beaker.
Add one MOLYBDATE HR No. 1 tablet straight from the foil to the
water sample and crush the tablet using a clean stirring rod.
Add one MOLYBDATE HR No. 2 tablet straight from the foil to the
same water sample and crush the tablet using a clean stirring rod.
Dissolve the tablets using a clean stirring rod.
Rinse out the vial with the prepared water sample and then fill to the
10 ml mark.
Close the vial tightly with the cap.
Place the vial in the sample chamber making sure that the
are aligned.
Zero
Press the [ZERO/TEST] key.
Test
Mo 3
The method symbol flashes for approx. 3 seconds.
Tablet
Powder
RESULT
The result is shown in the display in mg/l Molybdenum.
Tablet
Powder
Tolerance:
>
Conversion:
mg/l MoO
mg/l Na
12
All manuals and user guides at all-guides.com
0
15 mg/l: ± 1.0 mg/l Mo
15
30 mg/l: ± 1.5 mg/l Mo
= mg/l Mo x 1.67
4
MoO
= mg/l Mo x 2.15
2
6
marks
SC450_V1_04/2011

Publicidad

Tabla de contenido
loading

Tabla de contenido