2 System Operation

2.1 Instrument Check

A color check on green is recommended every 7 days, calibration on white is
forced every 30 days.
Check instrument on color tile
1.
Select Check in the main menu.
2.
Place unit on green color tile and tip on OK or press button Operate.
Values are within tolerance
1.
Confirm success message with OK.
2.
The unit returns to the main menu and readings can be taken.
Values are out of tolerance
1.
Confirm error message with OK and repeat green color tile reading.
2.
If the color check is not successful for three times, you are asked to per-
form a calibration on white tile and thereafter, a color check on green tile
again.
If calibration fails, clean the white tile. If calibration still fails, contact customer
support.