Preparing Instruments; Loading Rack - Amsco 7053L Manual Del Operador

Rack de tres niveles para instrumentos oftálmicos con complemento superior extraíble
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

2.3 Preparing
Instruments

2.4 Loading Rack

WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: Always load or unload
the Three-level Ophthalmic Rack
with Removable Upper Insert –
Long on appropriate loading cart
or surface.
Loading Instructions
To prepare instruments for processing, proceed as follows:
Disassemble all instrument components prior to loading them in
the Three-level Ophthalmic Rack with Removable Upper Insert –
Long and ophthalmic baskets.
Fully open all handles, jaws and graspers.
Manually clean any visible soil on cannulated instruments. Follow
device manufacturer's instructions for use to manually clean
instruments.
Pre-load ophthalmic baskets with the appropriate instruments.
Load a cataract ophthalmic instrument set in an ophthalmic
basket. Place basket on ophthalmic rack lower levels.
Load a vitrectomy ophthalmic instrument set in an ophthalmic
basket Place basket on the upper level on ophthalmic insert.
Cannulated ophthalmic instruments and suction tubes are
loaded onto special instrument holders on the ophthalmic insert.
Lower Part (Base Rack)
To load the base rack, proceed as follows:
1. Mount ophthalmic insert onto base rack.
2. Pre-load ophthalmic baskets with appropriate instruments.
Cataract ophthalmic instrument set is loaded on base rack.
Figure 2-5. Ophthalmic Basket Pre-Loaded
IMPORTANT:
Baskets
instruments to ensure cleaning efficacy.
Operator Manual
should
not
be
overloaded
with
2-5
10095339