Feellife A5 Manual De Usuario página 15

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Warning:
The masks must not be placed in hot water!
2. Disinfection
After each use, it is necessary to disinfect the components, including medication cup, lid,
connector, mask, etc. The disinfection methods are as follows:
1) Disinfection with hydrogen peroxide
Disinfect all the components by placing them in 2% hydrogen peroxide for 10 minutes, including
medication cup, lid, connector, mask. After disinfection, rinse all the parts with distilled water,
then wipe with clean and sterile gauze.
Do not use strong oxidizing agents such as perchlorate or disinfectants that are corrosive to
metals, polymer compounds.
Note:
1) Disinfectants remaining on components need to be wiped with sterile medical gauze to
ensure safe use next time. Do not touch the central area of the mesh diaphragm when washing
or cleaning the spray connector, so as to avoid damaging the mesh diaphragm.
2) The spray assembly is composed of a medication cup and a connector. When cleaning and
disinfection, you should loosen the connector by using a screwdriver.
3. Drying
Dry the parts carefully using a soft cloth. Shake the spray assembly gently(5~10 times), so that
the water inside the mesh is removed from the tiny holes.
Place the parts on a clean, dry and absorbent surface and leave them to dry completely (at
least 4 hours).
Note:
Please ensure that the parts are completely dry after cleaning, otherwise the risk of bacterial
growth is increased. Place the parts in a dry, sealed container. Ensure that the spray assembly
is completely dried off by shaking. Otherwise, the nebulization may not work after reassembling
the device.
15

Publicidad

Tabla de contenido
loading

Productos relacionados para Feellife A5

Tabla de contenido