Filled wrong medication types
Medication chamber is not completely
attached.
No medication in medication chamber.
After turning
power on, it is
immediately
No contact between medication and
power off.
mesh.
Electrodes on nebulizer and
medication chamber are stained.
Power is low.
Low battery power.
Batteries damaged.
Incorrect connection of AC adapter to
It is not
nebulizer.
nebulizing.
Rupture of mesh of medication
chamber.
Holes on vibrating mesh are clogged
with medication or water.
Vibrating Mesh of medication chamber
is severely clogged.
Drop medication and change appropriate
medication, or consult with physician
Re-attach the medication chamber correctly
and restart the power.
Put in the medication in the medication
chamber.
Adjust the nebulizer's angle so the
medication can come in contact.
Remove the stains and restart the power or
return to repair.
Charge the batteries by professional.
Charge the batteries and restart the power.
Otherwise need to replace new battery
group by professional.
Replace with new batteries by
professional.
Re-connect in the correct manner and
restart the power.
Replace with a new medication chamber
and then put in the medication.
Clear the holes of clogged medication
or water and restart the power.
If it still cannot be used after cleaning,
please return to repair.