9. Storage Conditions
The storage and transportation conditions of the products are detailed in the Product Technical
Parameters section.
· Keep the device out of the reach of unsupervised infants and children. The device may contain
small parts that can be swallowed.
· Prevent pets and pests from damaging equipment.
· Dry the parts immediately after cleaning and disinfection. Store the device and the
components in the environment that meets the requirements, be careful to avoid collisions.
· Direct sunlight, lint, dust may cause vibrating mesh rusted and oxidized and decrease
nebulization rate.
10. Trouble Shooting
Please refer to the table below to troubleshoot any problems you may encounter when using the
nebulizer.
problems
Medication chamber is not completely
attached.
No contact between medication and
Vibrating mesh more than 10 seconds.
Extremely low
Mesh of medication chamber is
nebulization.
clogged.
Vibrating mesh on medication cup are
clogged with medication or water.
Electrodes on nebulizer and
medication chamber are stained.
Possible cause
Re-attach the medication chamber correctly
and restart the power.
Adjust the nebulizer's angle so the
medication can come in contact.
Clean the medication chamber and
vibrating mesh. If it still cannot be used after
cleaning, please check vibrating mesh is
broken or not.
Clear the vibrating mesh of clogged
medication or water and restart the power.
Remove the stains and restart the power, or
return to repair.
16
solution