SET UP AND ASSEMBLY

Attaching the Ameda Platinum Breast Pump to Trolley

Trolley is available as an accessory item (Stock No. 17811).
• Attach pump to trolley prior to plugging cord into wall outlet.
• Align front of the Ameda Platinum Breast Pump with wording, "Front of Ameda Platinum and Elite™ Pump"
found on top of trolley mounting plate.
• Lower pump onto plate, inserting pump's four feet into holes, one in each corner of trolley mounting plate.
• Locate mounting hole with wording, "Ameda Platinum and Elite™ Pump Mounting" found on underside of trolley
mounting plate.
• Use mounting screw for the Ameda Platinum Breast Pump supplied with trolley to secure pump to trolley base.
Trolley Operating Instructions
• Once pump is attached to trolley, ensure all wheels are unlocked to move the trolley-pump assembly. Wheels are
unlocked by lifting the locking tab on each wheel.
• Move the trolley-pump assembly by gently pushing horizontally on the pump.
• When trolley-pump assembly is in the desired position, lock all wheels by pushing down on the locking tab on
each wheel.

Before Every Use

Always wash hands well with soap and water before handling the Ameda Platinum Breast Pump and

Ameda Pumpset.

IMPORTANT NOTE: ONLY USE THE AMEDA PUMPSET ON AMEDA ELECTRIC BREAST PUMPS.
Prior to starting pump, assemble and connect the Ameda Pumpset, per included instructions. If the Ameda Pumpset
is sterile, use as is. If not sterile, see the Ameda Pumpset cleaning instructions.
Ameda Pumpset
5

Single and Dual Pumping

The pumpset can be used to single or dual pump.
To single pump:
Remove one tube and
close tubing adapter.
8
|
ENGLISH
Adapter Cap
2
Silicone Diaphragm
1
Breast Flange
Valve
3
4
Milk Bottle
Tubing
Tubing Adapter

Ameda Platinum Breast Pump

To dual pump:
Both tubes should be
attached to tubing adapter.
(if applicable)
• Place pump on a horizontal flat
surface before use.
• Attach Ameda supplied grounded
power cord (available separately)
to AC power inlet on rear of
pump.
• Connect grounded plug to wall
outlet.
• Ensure pump is positioned so
that the power cord from the
rear is accessible for removal.
• Insert tubing adapter into
tubing adapter port.

Publicidad