Chlorine PP / M110
Chemical Method
DPD
Appendix
Interferences
EN
Persistant Interferences
•
All oxidising agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
•
Interference from Copper and Iron (III) are eliminated by the addition of EDTA.
•
Concentrations above 2 mg/l Chlorine, in the event of using Powder Packs, can lead
to results within the measuring range of up to 0 mg/l. In this case, the sample must be
diluted with chlorine-free water. 10 ml of the diluted sample should be mixed with the
reagent and the measurement taken again (plausibility test).
Interference
CrO
2-
4
MnO
2
Method Validation
Limit of Detection
Limit of Quantification
End of Measuring Range
Sensitivity
Confidence Intervall
Standard Deviation
Variation Coefficient
Conformity
EN ISO 7393-2
a)
determination of free, combined and total
10
from / [mg/l]
0,01
0,01
0.01 mg/l
0.03 mg/l
2 mg/l
1.68 mg/l / Abs
0.033 mg/l
0.014 mg/l
1.34 %
EN Method Reference Book 1.0