Er7
TANK HEATING
Er8
BOILER FILLING TIME-OUT
(ATMOSPHERIC MACHINE)
Er9
BOILER TEMPERATURE
ABOVE 100° C
If the boiler does not reach the set temperature after 40 minutes, the Er7
error message flashes on the TANK display (7) and the START key (3)
flashes red. All the functions remain active. To reset the alarm, switch off
the machine by pressing the ON/OFF key (1) for 3 seconds.
N.B.: Check the correct functioning of the tank resistance before switching it
back on again.
If the boiler has not been filled after 5 minutes, the filling valve switches
itself off. The Er8 error message flashes on the TANK display (7) and the
START key (3) flashes red.
To reset the alarm, press the START key (3) for 3 seconds to restart the
filling, or switch off the machine by pressing the ON/OFF key (1).
N.B.: Check and fix the cause of the filling failure before restarting the
machine. To restart the machine, press the ON/OFF key (1) for about 2
seconds
If the boiler probe exceeds 98°C, the machine switches itself off and all the
loads are deactivated. The Er9 error message flashes on the TANK display
(7) and the ON/OFF key (2) and START key (3) flash red.
To reset the alarm, switch off the machine by pressing the ON/OFF key (1).
N.B.: Check and fix the cause of the problem before switching the machine
back on. To restart, press the ON/OFF key (1) for about 2 seconds