Optima Auto Series / English
9. Trouble Shooting
9.1. User tackle-able issues
Issues that users might encounter and solve by themselves. See below for the description,
corresponding display and solutions.
(1). Low Pressure Alarm
Alarm will be indicated when the pump detects low pressure condition within the mattress.
NOTE- In alternating mode, the system first inflates and then has the cells hold at a certain
level to support the patient while monitoring the cells at the same time. The situation
continues until the low pressure alarm is lifted.
NOTE- The pump seals the air inside the mattress, which temporaily keeps the patients from
bottoming out during low pressure condition. Patient should be moved if the power failure
issue cannot be solved soon.
(2). Low battery capacity
NOTE- When low battery symbol is shown on the display, plug the power cord into the wall
outlet immediately.
CAUTION- Pump system will stop running;
buzzer images when the battery capacity is blow 10%.
power cord into the wall outlet, and press the power key to restart the pump.
Code No.
01, 02, 05
03, 04, 06
07
19
Diagram
Low pressure alarm:
Low pressure alarm: foot
Low pressure alarm:
Position sensor is not
the LCD displays the alarm, along with the
To restore the system, plug the
Description
body section
section
head section
connected
What to Do
(1). Check all the joint
parts to see if they are
well connected, or if
there is leakage, along
with kinked or broken
parts that need to be
addressed.
(2). Call for service if the
problem persists.
Re-start the pump to fix
this. Call for service if the
problem persists.