Descargar Imprimir esta página

VWR avantor OXY 11 Manual De Instrucciones página 12

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 29
EN
English
Thoroughly rinse the sensor
head with deionized water.
Fill a new membrane cap with
ELY/G electrolyte solution.
Ventilation area
points upwards
Hold the sensor inclined and
screw on the membrane head
fingertight using a paper
towel. Excess electrolyte
solution is forced out of the
ventilation area.
12
H O
2
Water the counter electrode in
deionized water for at least
10 minutes.
Remove any air bubbles by
carefully tapping the
membrane head. Additionally,
you can prevent air bubbles
by throwing the first filling
away and refilling the
membrane cap.
3 mm
Check the filling:
Inspect the face surface. No
air bubbles may be present
within the dashed circle. Air
bubbles outside this area do
not interfere.
Carefully shake off the drops
of water.
Rinse the sensor head with
electrolyte solution.
ba75842e04
OXY 11
05/2021

Publicidad

loading