Change the operation method
< Basic operation >
When the manikin is turned on, the manikin always operates in <basic operation>, and
when chest compressions with a constant depth of compression over 3cm for 5 seconds
are performed, it will switch to <specialized operation>.
The 'normal circulation' indication light will illuminate when chest compressions with
a depth of compression over 5cm and a speed between 100 times and 120 times per
minute in <basic operation> or chest compressions with a depth of compression
between 5cm and 6cm and a speed between 100 times and 120 times per minute in
<specialized operation> are performed.
▒
Note
When the manikin is turned on, a one-time automatic calibration will be carried out for approximately 2 seconds.
It is recommended that chest compressions on the manikin not be performed when automatic calibration is being carried out.
If the chest compression is not detected for 2 seconds, the blood circulation indication lights, the compression depth indication lights and the normal circulation indication light will
be turned ou and the manikin will enter stand-by mode. When in stand-by mode, the power status indicator lamp will flash green.
When the user carries out chest compressions again while the manikin is in stand-by mode, the chest compressions will be detected automatically and the indicators will operate
again. However, if left in stand-by mode for 30 minutes, the manikin will switch itself ou to conserve the batteries. To switch the manikin on again, slide the power switch to the ' ou '
position for more than 2 seconds and then switch it on again.
While in stand-by mode, the power consumption is very low. However, we recommend that you switch ou the manikin when not in use.
If the battery is low, the power status indicator lamp will flash red. At this time, replace the battery with a new one.
< Specialized operation >
▒ Change the clicker setting
< ON >
Turn the 'clicker' on and ou using the switch as shown. The clicker makes a "clicking"
sound when the chest is compressed more than 5cm.
< OFF >
5