Guidance and manufacturer's declaration - electromagnetic Immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of
the Device should assure that it is used in such an environment
Immunity Test
Electrostatic
±8 kV contact
discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast tran-
Power supply lines:
sient/burst
±2 kV
IEC 61000-4-4
input/output lines:
±1 kV
Surge
line(s) to line(s):
IEC 61000-4-5
±1 kV.
line(s) to earth:
±2 kV.
100 kHz repetition
frequency
Voltage dips, short
5% UT
interruptions and
(>95% dip in UT)
voltage variations on
for 0.5 cycle
power supply input
lines
40% UT
IEC 61000-4-11
(60% dip in UT)
for 5 cycle
70% UT
(30% dip in UT)
for 25 cycle
<5% UT
(>95% dip in UT)
for 5s
Power frequency
30 A/m
(50/60Hz)
50Hz/60Hz
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
16
.............................................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL MQ5800, MQ5900, MQ5900P
.............................................................................................................................
IEC 60601
Compliance level
Test level
±8 kV contact
±15 kV air
Power supply lines:
±2 kV
line(s) to line(s): ±1 kV.
100 kHz repetition
frequency
5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycle
70% UT
(30% dip in UT)
for 25 cycle
<5% UT
(>95% dip in UT)
for 5s
30 A/m
50Hz/60Hz
Electromagnetic environment - gui-
dance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Power frequency magnetic fields should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environment.