When lifting or lowering a patient, it is recommended to press and hold the relevant button. Repeatedly
pressing and releasing the buttons will cause the actuator motor to start and stop each time, using more battery
and shortening its operating lifetime.
The electrical system includes a emergency stop red button. When the red button is in the out position, the
battery will operate normally. When the red button is depressed, the power from the battery is cut off. To reset
the button to the out position, turn it clockwise and release.
The knee´s support (see page 5) is adjustable: allowing five different heights (50, 54, 58, 62 and 66 cm from
floor.
This support is basculable to guarantee the best support for knees adapting to each patient.
It has a strap with safety lock in case it is necessary to adjust the legs of the patient
The boom of the crane is telescopic (see page 4) to adapt to different measurements (4 positions every 30 mm).
To SHORTEN or EXTEND the boom, release the safety button (L) and slide the mobile boom towards the mast to
SHORTED and in the opposite direction to EXTEND.
4.2.- Recharging battery
System has a noisy system when the battery is near to finish, with this is possible made complete cycles.
If you only have one removable battery and one charging base, place the hoist near an electric socket, connect
the power supply cable to the charging base and plug into the mains (110-220V). (see page 4-H)
While the battery is charging the actuator fails, the emergency stop should not be pressed and green LEDs "on"
lights. The battery indicator lines flash. When fully charged the 4 battery indicator lines will remain fixed The green
light "on" will remain on until you disconnect from the net-work.
If the hoist is used daily, it is recommended to recharge the battery during the night. The charger will not
allow the battery to overcharge.
When not using the hoist, it is recommended to re-charge the battery at least once every three months to
ensure it is always in good working condition.
Do not disconnect the black charger jack plug by pulling the cable. When not recharging the battery, it is
recommended to keep the mains cable plugged into the charging base to prevent any liquids accidentally
entering the socket.
If you have an extra charging base (on request), you can remove the battery from the charging base mounted
on the hoist and charge it using the extra charging base, suitably located near a mains electricity socket.
4.3.- Changing the battery
To remove the battery from the charging base, hold the upper handle, press and remove carefully of its support.
To place the battery back in the charging base, slide it on position, join the upper handle on the support until