Setup and connection
50
8.4 Connecting the vacuum controller to the
vacuum system
Connecting the vacuum controller via Bluetooth
Plug a Bluetooth stick into the rear of the vacuum controller.
To do this, remove the cover on the rear of the vacuum con-
troller. Then you can plug the Bluetooth stick in and place the
cover in position again (see Fig. 10).
Plug the other Bluetooth stick into the connector provided for
this purpose on the vacuum system (see Fig. 23/2).
The Bluetooth sticks are designed specifically for the
vacuum controller and the vacuum system and must not
be mixed up. For this reason they are placed in labeled
pouches. When using them, ensure that the correct
sticks are plugged into the vacuum controller and the
vacuum system respectively.
A PC can be connected via a mini-USB port in parallel
with the Bluetooth connection.
Fig.10: Plugging Bluetooth stick into vacuum controller
Connecting the vacuum controller via cable
Connect a USB A cable to the vacuum controller. This can be
connected to the vacuum system via the mini-USB port.
When the vacuum controller is connected to the vacuum
system with a cable, the mini-USB port cannot be used
to connect a PC. To connect PC, the USB cable be-
tween the vacuum controller and the vacuum system
must be removed.
Translation of Originial Operating Instruction, KNF 339416-339418 02/23