H i k o n e b | 9 0 8 D C
Us e r Ma n ua l
8. Troubleshooting
If a fault occurs, first switch off the device. To do this, press the power button. If possible,
eliminate the cause of the fault and switch the device on again.
Error
The device is not
operational,
On-Off button light is
not on
On-Off button is on
but there is no aerosol
production
Red error LED is on
Aerosol production is
visible, but not coming
out of tubing
8a. Replacement of the main fuse.
WARNING!
Electric shock! Before replacing the fuses, disconnect the device from the mains.
The main fuse (2A) is located inside the IEC socket on the back of the device. Remove
the power cord from the device. To remove fuse holder, use a small flat screwdriver
pulling the holder forward. Replace fuse with the same fuse rating and replace fuse
holder by pushing back into position.
26
Reason
No mains
1. Transducer
2. O/C fuse
Low Water Level
Nebulizing chamber not locked
on the device correctly
Blockage of patient tube due to
wrong positioning
Occluded Filter
Flow menu disabled
Faulty fan
Solution
Check plug, IEC connections and
fuses
1.Replace transducer
2. Replace fuses
1. Add more water
2. Re-fit chamber ensuring correct
Change position of the aerosol tube
to point upwards and straight
Replace the filter
Activate Flow setting
Replace Fan