Chromium-plated
Nut, transducer
installed
NOTE!
The device will not operate if the transducer is not installed correctly. Also, whenever the
transducer has been replaced and before the chamber is connected in place, fill the
chamber with water and check for possible leaks.
10. Cleaning, Disinfection, Sterilization
Cleaning
The device and 5 feet mobile stand can be cleaned with a mild detergent and damp
cloth. For special cleaning/disinfecting please follow the manufacturers cleaning
instructions detailed in this user manual.
NO parts should be exposed to excessive pressure during the cleaning process.
Pay specific attention to the transducer surface ensuring it is clean. Clean lime scale
residue by wiping, not scrubbing. Abrasive scrubbing can damage the transducer.
Disinfection
When disinfecting the device surface, the device's surface, make sure that the mains
power cable is removed. Ensure that no liquid enters inside the device when cleaning.
Operate the device only after the disinfection has completely dried and evaporated. Use
only disinfection products consisting of aldehyde, ammonia and alcohol components.
Never use phenol derivative disinfections, as this will compromise and reduce the
service life of plastic parts.
28
Transducer
Chromium-plated
nut