CAUTION To avoid damage to the product, observe the following general precautions when disinfecting.
Do NOT use Povodine
Do NOT use strong oxidants such as bleach.
Do NOT use bleaches containing sodium hypochlorite.
Do NOT use disinfectants containing iodine complexes.
If you intend to use a cleaning agent not listed here, check its material compatibility first.
ALWAYS dilute according to the manufacturer's instructions.
Recommended Disinfecting Substances:
Aldehyde-based
Alcohol-based
Sterilizing
CAUTION Do not use bleaches containing sodium hypochlorite (for example Clorox™) on any of the cables.
Patient Module M1364A, Adapter Cable M1362B
Never sterilize by means of autoclaving, gas treatment, formaldehyde process, or radiation.
Adapter Cables M1362A, M1363A
Only ECG adapter cables M1362A (no longer available) and M1363A can be sterilized. They can be
sterilized by means of autoclaving or by gas sterilization. We recommend that you sterilize after cleaning,
and only when necessary as determined by your hospital's policy. The M1362A and M1363A cables have
been tested to withstand Ethylene Oxide (EtO) gas sterilization. Be sure that all safety precautions regarding
aeration after EtO exposure are followed. The cables are tested to withstand autoclaving at 136ºC (277ºF)
maximum.
M1364A Specifications
Overall length: 2706mm (+30, -100mm) / 106.5in (+1.2, -3.9in)
Length of free cable: 2618mm (+30, -100mm) / 103.1in (+1.2, -3.9in)
Weight: 120 grams / 4.2 ounces
Size: 88x42x30mm / 3.5x1.7x1.2in
Red Socket: DECG or MECG connection
Water Ingress Protection Code: IP 67 according to IEC 60529 (2001)
(dust tight, immersible in water to a depth of 0.5m for 30 min.)
Resulting Isolation (DECG or MECG mesurement): CF
Disposal
Dispose of the equipment in accordance with your country's laws.
WARNING
To avoid contaminating or infecting personnel, the service environment or other equipment, make sure
the equipment has been appropriately disinfected and decontaminated before disposal at the end of its
useful life.
4
®
®
®
, Sagrotan
, Mucovit
or strong solvents.
®
Cidex
Ethanol 70%, Isopropanol 70%