Descargar Imprimir esta página

Agilent IDP-10 Manual Del Usuario página 173

Publicidad

Instructions for Use
Start-up and Operation of the IDP-10 Dry Scroll
Vacuum Pump
Startup Procedure
Check that the inlet screen is installed before beginning operation.
Do not insert a finger or any foreign object in the path of the fan; serious
WARNING
personal injury may result or the pump may be damaged.
1
Operate the pump at an ambient temperature of 5 °C to 40 °C (41 °F to 104 °F),
otherwise damage to the pump or shortened operating life may result.
Do not block the fan ducts. Blocking these ducts can cause pump overheating.
CAUTION
A pump surface temperature in excess of 65 °C (150 °F) is potentially damaging.
If such conditions are observed, turn the pump off and allow it to cool.
Disassemble, inspect for damage, and repair, if necessary.
2
Make sure that the pump is configured for the mains voltage to which the
pump is connected.
The pump ON/OFF switch is a slide type switch that has symbols in accordance
NOTE
with IEC Publication 417 to represent the ON and OFF positions. The following
figure shows a switch in the ON position.
Agilent IDP-10 Dry Scroll Vacuum Pump User Manual
173

Publicidad

loading