Inrush Current; Mechanical Connections; Isolation Valve; Pump Inlet - Agilent IDP-10 Manual Del Usuario

Tabla de contenido

Publicidad

Technical Information

Inrush Current

The following table lists the typical IDP-10 maximum starting current and duration.
Table 17 Typical IDP-10 Maximum Inrush Current (A)
Frequency

Mechanical Connections

Isolation Valve

The IDP-10 has an optional integral automatic isolation valve. If the IDP-10 is
equipped with this valve, an internal timer opens the valve 30 seconds after the
pump is switched ON. If power is lost or the pump is switched OFF, the isolation
valve will immediately close.

Pump Inlet

Use NW25, or larger, clean vacuum hardware with as short a length as practical
between the pump and vacuum chamber.
Insert a bellows between the pump and vacuum chamber to provide both vibration
isolation and strain relief.

Pump Exhaust

A female 1/4" National Pipe Thread exhaust fitting is located radially near the front
of the pump. Additionally, an NW16 male adapter with 1/4" National Pipe Thread is
provided. To avoid overheating the pump, do not restrict the exhaust flow with long
lengths of small diameter tubing. Use as short as practical lengths of NW16, or
larger, diameter hardware
Agilent IDP-10 Dry Scroll Vacuum Pump User Manual
100 V
60 Hz
49
50 Hz
50
57 Hz
N/A
63Hz
N/A
High Voltage Range
120 V
127 V
200 V
50
50
25
N/A
N/A
49
N/A
50
N/A
N/A
50
N/A
230 V
240 V
N/A
38
32
35
N/A
N/A
N/A
N/A
191