LAMP MODEL:
TECHNICAL DETAILS
Illumination E
at 1 m ± 10% [Lux]
c
Illumination E
at 0.7 m ± 10% [Lux] (DENTAL CARE)
c
Colour temperature [K]
Colour rendering index R
R
[-]
9
Light field diameter d
50
Light field diameter d
10
Light field diameter [mm]
Light field diameter [mm] (DENTAL CARE)
Lighting depth L1+L2 [cm] (60%)
Lighting depth L1+L2 [cm] (20%)
Primary alternate voltage [Volt ac]
Frequency [Hz]
Absorbed power [VA]
Light source
Duration of LED diode light source [hr]
(this figure can vary according to power peaks and
operating frequency)
Light intensity control [%]
Colour
Directive
Standards
Classification of Medical Device
Distribution of minimum and adequate lighting (luminous flux emitted by the EM equipment
shall not vary by more than 20% during use; the colour temperature and colour rendering
index shall be stable and within the range 3000K-6700K and 85-100, respectively; E
Essential
performance
Limitation of energy in the operating field (UV-irradiance for wavelengths below 400 nm
shall not exceed 10 W/m
1000 W/m
Operation and
maintenance manual
8 Technical properties
[-]
a
[mm]
[mm]
value shall be ≥ 40,000 lux and ≤ 160,000 lux).
2
; and the total irradiance E
at a distance of 1000 mm; Ec value shall be ≥ 40,000 lux and ≤ 160,000 lux;
2
MU_029_EN
Rev.0
ORION TECH
4,500
20 - 100
93/42/EEC and 2007/47/EC
EN/IEC 60601-2-41
in the lighted area shall not exceed
e
E
/E
≤ 6 mV/m
2
lx).
e
c
07/11/18
EN
Page 24 of 35
VALUE
130,000
70,000
5,000
96
≥ 90
95
190
270
200 x 150
27 + 13
55 + 40
100 - 240
50 / 60
60
29 LED
> 60,000
20 - 100
RAL 9003
Class I
c