no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils
radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1)
l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout
brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le
onctionnement.
INTERNAL BATTERY INFORMATION
This Normatec 3 control unit is equipped with a rechargeable lithium ion battery. The internal
battery is designed to allow use of the Normatec 3 System anywhere—even when power
outlets aren't available. The Normatec 3 control unit may need to be plugged in before
first use. A fully charged battery will provide power for 2+ hours of continuous use. It takes
approximately 6 hours to fully charge the battery when the control unit is plugged in and not in
use. The rechargeable lithium ion battery is intended to be changed only by authorized service
personnel with the use of a special service tool.
PRODUCT SPECIFICATIONS
• Normatec 3 Model: REJ6
• Normatec 3 Dimensions: 4" (width), 4" (depth), 8.5" (height); [10.2 cm (width), 10.2 cm
(depth), 21.6 cm (height)]
• Normatec 3 Weight: 3.2 lbs [1.45 kg]
• Normatec 3 electrical requirement: 15V
• Maximum Air Pressure: 110 mm Hg
• Temperature (operating): +41° F to 104° F [+5° C to +40° C]
• Temperature (storage): -13° F to +158° F [-25° C to +70° C]
• Relative Humidity (operating): 15% to 93%, non-condensing
• Relative Humidity (storage): -25˚ C without relative humidity control; +70˚ C at relative
humidity up to 93%, non-condensing
• Atmospheric pressure (storage and transportation): 190hPa to 1060hPa
• Atmospheric pressure (operating): 700hPa to 1060hPa
AC-DC ADAPTER
WARNING! Only use the AC-DC adapter model number 30120 provided with the system.
Using a different adapter may cause the system to not operate correctly.
• Input: 100-240V 0.8-0.4 A 50/60 Hz per Normatec 3 model number 60090-001-00
• Output: 15V
ELECTROMAGNETIC COMPATIBILITY
The information contained in this section (such as separation distances) is in general
specifically written with regard to the Normatec 3. The numbers provided will not guarantee
faultless operation but should provide reasonable assurance of such. This information may
not be applicable to other medical electrical equipment; older equipment may be particularly
susceptible to interference.
GENERAL NOTES
Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC
information provided in this document and the remainder of the instructions for use of this
control unit. Portable and mobile RF communications equipment can affect medical electrical
equipment. Cables and accessories not specified within the instructions for use are not
authorized. Using other cables and/or accessories may adversely impact safety, performance,
and electromagnetic compatibility (increased emission and decreased immunity). Care should
be taken if the equipment is used adjacent to or stacked with other equipment; if adjacent or
stacked use is inevitable, the equipment should be observed to verify normal operation in the
configuration in which it will be used.
DC minimum 1.6 A per Normatec 3 model number 60090-001-00
DC 1.6 A