The grinding chamber interior (funnels, dosing equipment,
rotor, grinding surface, filter and outlet nozzle) with a brush
(no wire brush) should be cleaned after the grinding proce-
dure. You can also use rags or cleaning brushes. The grind-
ing surface does not need to be disassembled for cleaning
between grinding operations with the same feed material. If
a new grinding procedure with other feedings is to be per-
formed whereby no contamination may be produced by the
previously ground material, then more thorough cleaning is
necessary which means dismantling the grinding chamber
and the rotors. See the chapter „Commissioning" for assem-
bly and disassembly.
The grinding chamber parts can be washed with water and,
if required, with a conventional dish-washing agent.
Be certain that the parts are dry before being reassembled.
Caution: Refer to the chapter on „Safety Instructions"
when cleaning the cutting-grinding surface and the blade
rotor as the screwed on blades are very sharp and may result
in injuries if they are improperly handled.
Clogged filters can be cleaned with brushes or compressed
air. If ground material parts have settled into the filter holes
in such a way that cleaning alternatives are not sufficient,
then you will have to clean the clogged holes with suitable
needles or clean the filter in an ultrasonic bath.
If beaters or blades and the grinding surface become worn
in the course of the time, particularly with constant grinding
of very hard substances, so that quick operation with the
appropriate quality is no longer possible, then the worn out
parts should be replaced.
Blades may not be reground as the blade clearance
will become too large. Fore reasons of safety, only
new, original IKA bolts may be used for fastening
the knives.
Sieve insert with hole sizes:
Ø 0.25 mm
Ø 0.50 mm
Ø 1.00 mm
Ø 2.00 mm
Ø 3.00 mm

Maintenance and cleaning

The mill IKA MF 10 is maintenance-free. It is subject only to
the natural wear and tear of components and their statistical
failure rate.
Cleaning:
Use only cleaning agents which have been approved by IKA to
clean the device: Water containing surfactant / isopropyl alcohol.
- Wear protective gloves while cleaning the devices.
- Electrical devices may not be placed in the cleansing agent
for the purpose of cleaning.
- Do not allow moisture to get into the device when cleaning.
- If a different cleaning or decontamination method than the
method defined by IKA is planned, the user must ascertain
with IKA that this method does not damage the device.
Spare parts order:
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item and designation of the spare part,
see www.ika.com, spare parts diagram and spare parts list.
Repair:
In case of repair the device has to be cleaned and
free from any materials which may constitute a
health hazard.
For repair, please request the "Decontamination Certificate"
form IKA or use the download printout of it from IKA website:
www.ika.com.
If you require servicing, return the appliance in its original
packaging. Storage packaging is not sufficient. Please also
use suitable transport packaging.