1 . To turn the thermometer on, push the "start" button
as shown on page 3 .
During an internal self-check, the display
shows all segments .
The reading shown on the screen is the last
temperature taken .
Flashing symbol of lens filter will appear if no
lens filter is attached . Temperature can be
taken only if lens filter is mounted on
the probe .
The thermometer is ready to use after
two beeps and the non-flashing ear
icon appears .
2 . Fit the probe (3) snugly into the ear canal,
then push the "start" button (4) .
One beep signals normal temperature . Audio
temperature indicator will trigger 2
consecutive beeps when the reading is equal
or higher to 99 .5 °F (37 .5 °C) and 4 beeps if the
reading is equal or higher to 101 .4 °F (38 .6 °C) .
The result is shown on the display (7) .
3 . Gently pull off the hygienic lens filter (2) .
Do not use force in removing the hygienic
lens filter .
4 . Put on a new, clean, hygienic lens filter by
snapping into position .
5 . Place the protective cap on (1) .
The Braun ThermoScan® 4 Ear thermometer turns
off automatically after 60 seconds of inactivity .
The thermometer can be turned off by pressing
and holding the start button for more than
5 seconds .