Storage and transport

WARNING
Risk of equipment damage
The measurement gas dampening mechanism must be
removed before the Air Guard is disassembled, and the
connectors must be sealed with the protective caps included in
delivery.
► Deinstallation and re-commissioning of the Air Guard may
only be done by authorised technical personnel.
Close the inlet connections with dust caps to prevent
contamination. Observe the temperature range (see chapter
9). Protect the product against direct sunlight.
Store the device in the supplied original packaging during
longer times of storage or transport.
8

Disposal

The product must be disposed of according to the national
waste disposal regulations or by an appropriate waste disposal
company.
8.1
Disposal of electrical and electronic
equipment
Electric and electronic equipment are subject to
national and international waste disposal regulations
and must not be disposed of as residential waste. This
is indicated by the adjacent symbol. Further
information is available online at www.draeger.com.
Dräger accepts returns of waste electrical and electronic
equipment for recycling or disposal.
9

Technical data

Dimensions
(width x height x depth)
Weight
Air Guard 6500 with all
options
Air Guard 6700 with all
options
Temperature range
Operating voltage
Operating pressure
HP
LP
Relative humidity
Intake air and ambient air
Power intake
Volume of the audible
warning
760 x 600 x 210 mm
approx. 41 kg
approx. 42 kg
0 to +40 °C
120-230 V/AC
200 to 300 bar
6 to10 bar
At least 40 %
Free of ammonia or
ammonium compounds
2.2 to max. 4.4 A
83 dB (at a distance of 10 cm)
Data on potential-free contact:
Cable bushing
Maximum possible conductor
cross-section on contact
Maximum permissible
switching current
Maximum permissible
switching voltage
Sensor measuring range
CO
CO
2
O
2
VOC
Dew point
H
S (optional)
2
The measurement accuracy of the sensors can be found in the
relevant instructions for use for the sensors.
10

Order list

Designation
DrägerSensor CO
DrägerSensor O
LS
2
DrägerSensor H
S
2
50 ppm CO test gas
500 ppm CO
test gas
2
20.9 % O
test gas
2
5 ppm isobutylene test gas
15 ppm H
S test gas
2
99.999 % N
test gas
2
Dräger-Tubes activated carbon type G
(pack of 10) (for gas treatment)
Storage and transport
M 16x1.5 screw connection
(approx. 7 to 10 mm clamp/
seal zone)
2
2.5 mm
6 A
250 V/AC or 30 V/DC
0 to 50 ppm
0 to 2000 ppm
0 to 25 Vol%
0 to 1.6 mg/m³
0 to 5000 ppm
0 to 15 ppm
Order number
68 09 605
68 09 630
68 09 610
68 11 117
68 12 383
68 13 239
68 13 685
68 14 199
68 10 394
67 28 831