3. Maintenance
3.1 Troubleshooting
Fault
The scaling tip doesn't
The power plug is in loose contact.
vibrate and there is no
The foot switch is in loose contact.
The fuse in the adapter or main unit is broken.C
stepping on the foot switch.
The tip is in loose contact.
The scaling tip doesn't
The connect plug between the handpiece and the circuit board
vibrate but there is water
is in loose contact.
Something wrong with the handpiece.
on the switch.
Something wrong with the cable.C
The scaling tip vibrates
but there is no spray when
The water control knob is not on.T
stepping on the foot switch.
The tip hasn't been screwed on to the handpiece tightly.
The tip is loose because of vibration.
The vibration of the tip
becomes weak.
The coupling between the handpiece and the cable isn't dry.
The tip is damaged [note 2].
There is water seeping from
the coupling between the
The waterproof "O"ring is damaged.C
handpiece and cable.
Possible cause
Solutions
Make the plug insert to the socket well.
Insert the foot switch to its socket tightly.
ontact our dealers or us.
Screw the tip on the handpiece tightly (See Picture 4).
Contact our dealers or us.
Send the handpiece to our company to repair.
ontact our dealers or us.
urn on the water contrl knob [note 1].
Screw the tip on the handpiece tightly (See Picture 4).
Screw on the tip tightly (See Picture 4).
Dry it by the hot air.
Change a new one.
hange a new waterproof "O"ring.
9

Publicidad

loading