4.When the measurement is complete, the systolic and diastolic pressure
readings and pulse rate are displayed. The cuff exhausts the remaining
air and deflates completely.
mmHg Display
5.Press the [START/ STOP] button again to turn off the power. If no
operation for 50s, the device will turn off automatically.
SPECIAL CONDITIONS
CAUTION
Inflating to a higher pressure than necessary may result in bruising where
the cuff is wrapped by. If your systolic pressure is known to be more than
220mmHg, press and hold the START/STOP button until the monitor
inflates 30 to 40 mmHg higher than your expected systolic pressure
NOTES:
• Do not apply more pressure than necessary.
• The monitor will not inflate above 299mmHg.
• When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
• If it is necessary to interrupt the measuring for any reason (e.g the
patient feels unwell), press the Start/Stop button at any time.
SYS
SYS
DIA
DIA
PUL
PUL
/min
/min
kPa Display
12
Memory Function
Checking Memories
1.Press the [MEM] button to enter the memory function.
2.The first display with AVG is the average value of the latest 3
measurements, then the most recent measurement memory.
(M1, M2, M3...)
Note:
a.All the memories of the selected user will be consistent with your unit
setting.
b. If the measurement memory shows arrhythmia, there is a sign of
arrhythmia.
c. This device stores 2*90 sets of memories.
Deleting Memories:
First Press the [MEM] button and hold on, then press the [SET] button for
3 seconds, the screen displays (
measurement data is deleted.
Note:
This operation will delete all the recorded data.
Please keep the records in another way in case that you may need it later.
) to indicate that all the existing
13