Error Indicators & Troubleshooting Tips
SYMBOL
CAUSE
Arm plug disconnected
E-1
Arm cuff not applied correctly
Air is leaking from the arm cuff
E-2
Cuff over inflated
E-3
Inflation Failure
E-4
Movement during
measurement
E-5
Clothing is interfering with
the arm cuff
E-6
Device error
Batteries are worn out.
CORRECTION
Insert the plug securely
Apply the arm cuff correctly
Replace the cuff with the New one
Repeat measurement. Remain still and do
not talk during measurement.
If "E2" appears repeatedly , inflate the cuff
manually until it is 30 to 40 mmHg above your
previous measurement result.Refer to "Taking
a Measurement"
1.Check if the air plug is twisted or not.
2.Check if the air plug or the air jacket is
properly inserted in the device.
3.Check if the air plug or the cuff is broken
Repeat measurement. Remain still and do not
talk during measurement. Refer to "Taking a
Measurement".
Remove any clothing interfering with the arm
cuff. Refer to "Applying the Arm Cuff"
Contact Customer Service.
Replace the four batteries.Refer to"Battery
Installation"
14
TROUBLESHOOTING TIPS
PROBLEM
No power
No display appears on the unit.
Measurement values appear
too high or too low.
SYMBOL DESCRIPTIONS
The following symbols may appear in this manual, on the Digital Blood
Pressure Monitor AXD-808, or on it's accessories. Some of the symbols
represent standards and compliance associated with the Digital Blood
Pressure Monitor AXD-808 and its use.
CAUSES AND SOLUTIONS
Replace all four batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Blood pressure varies constantly. Many
factors including stress,time of day,and
how you wrap the cuff, may affect your
blood pressure. Review the sections
" Before Taking a Measurement "and
" Taking a Measurement ".