The life of a rechargeable battery depends on the number and type of cycles of charge
/ discharge. If WELLMAG is not used frequently, we recommend recharging the battery
monthly; to prolong the life of the battery we recommend once in a while discharging
it completely (see empty battery symbol) and to have a complete 8-12 hours recharge
cycle.
When the battery's life is significantly lower than it used to be, we recommend replacing
it. For this, please refer to an authorized Centre of Assistance that will replace it under
all security and safety requirements.
Never reverse the polarity of the connections of the battery, not to cause the destruction of the
1)
Do not throw the exhausted battery with the ordinary wastes. Please take them to
the corresponding collection site.
2)
Do not manipulate/ throw the battery into a fire.
3)
Do not short-circuit the terminals.
4)
Avoid sparks or flames, above or around the battery.
5)
In case of direct contact with liquid leaking from the battery, wash immediately
with water. In case of direct contact with eyes, wash abundantly with water and
consult immediately a doctor.
WellMag
Error message on
display
SOL
Unproper connection of the
solenoids into the device
OL
Over Load, problem in the
solenoids
18
CAP7 – Power Supply
7.3 - Battery maintenance
7.4 - Battery replacement
electronic circuits.
7.5 - Precautions for use
CAP8 - Error messages
Problem
Solution
Make sure the connector
is properly inserted to start
the therapy. If the problem
persists, contact the Technical
assistance
Contact the Centre of
Assistance
CAP9 – Maintenance
9.1 - Solenoids
The connection's cable must be periodically revised to check the possible presence
of cracks, possible cause of dispersion of the magnetic fields. Clean the solenoids
periodically with a damp cloth.
9.2 - Device and power supply
We recommended cleaning the device and power supply with a damp cloth. Do not
use in any liquid, because these devices are not liquid proof (IP20).
9.3 - Power supply replacement
Check periodically the level of wear of the isolating parts of the power supply before
connecting it. If it's damaged, also only partially, replace immediately the power
supply.
9.4 - Immediate maintenance
Immediate maintenance must be performed by authorized staff if one of the
followings occurs:
external mechanical stress (eg. serious falls);
strong overheating (eg. if the device has been left close to sources of intense heat);
some liquids may have been penetrating the device;
the power supply, or other parts of the device are damaged, broken or missing;
the functionality of the instrument seems altered.
Use the device with the endowed accessories only
Run every year a secur-tester to check the device following the legislation EN60601-1. The useful life of the
device is guaranteed only if such maintenance is regularly performed.
WellMag
19