WellMag
Pay attention and use carefully the device with solenoids, not to compromise the effectiveness of the treatment.
Make sure that the electric system is in conformity with the actual safety norms.
The instrument has degree of protection IP20 (see Chap. "Technical features"). Do not use it close to liquids,
because it's not liquid proof.
Do not use close to cellular telephones (maintain at least one meter of distance).
The stability of the emission can be affected when operating close (less than 1 meter) to an instrument with
short waves therapy or microwaves.
Do not use with the presence of devices to monitor the patient's vitals.
Do not connect simultaneously the patient with the WELLMAG and with a surgical instrument High Frequency,
to avoid dangers for the patient and for the instrument.
The device works accordingly to its specifications if the environment is maintained to a temperature
between the 5° and the 40° C and with humidity inferior to the 80%. The same conditions must be
maintained during transportation and the storage.
In case of malfunctioning, do not use and request service of reparation.
Do not operate in proximity of inflammable substances.
Use only gels and accessory that are endowed with the device.
It's important to inform the patient on the type of sensation they will experience during the therapy. In case
the patient perceive a different sensation, it is important to stop the treatment and to remove the solenoids.
If the intensity is causing overheating or pain in the treated zone, it is necessary to reduce immediately the
intensity of stimulation or eventually interrupt the treatment.
Keep away from children.
20
CAP10 – Warnings
CAP11 – Technical features
11.1 - Power supply
Power supply: Biron brand
PRI: 230V 50Hz SEC: 13,6V- 1,2A
Internal power source: Rechargeable battery Li-Ion 12V-3,8 Ah
11.2 - Output features
Max intensity of magnetic induction (B): 100 Gauss
Frequency (f): 1-200 Hz
11.3 - Other features
Dimensions: 140x105x40h [mm]
Weight: 400 [g]
Class: II Type: BF
Classification to liquids entry: IP20
Safety in presence of flammable anaesthetic gas: it's not of AP or APG category
Type of operation: continuous
S.N. WL MG ZZ XXXX
WellMag
Serial number (S.N)
WELLMAG identity code
Device identity number
Year of production
21