4. If calibration is successful, the LCD respectively displays "C-L/C-H",if not
successful, the LCD displays "no".
5. If the temperature still has some departure, you can repeat calibration
according with the above steps.
NOTE: Suggest using 191 or 192 thermometer to measure the temperature.
2.10.
2.10.1. Replacing Heating Element
1. Replace the heating element after the handle has cooled down.
2. Unscrew and remove the handle enclosure for preventing burned.
3. Unscrew and remove the three screws on fixing the steel tube, in which has
the heater element assembly. And then it can pull out the heater element
assembly to change a new one (the heater element assembly and the steel
tube are incorporate and don't need disassemble).
4. When installing the new heater element assembly, it must insert the pins of
the heater element assembly to the corresponding sockets, and insert the
sensor pins to the sensors sockets.
5. Reassemble the handle according to the opposite order of disassembling. The
heater element must be insert into the end.
CAUTION:
Replacing the parts
Avoid invalidate the ground cord when replacing the
heater element.