Patient lift cannot work (ascend and descend)
Charger cannot charge the battery
Patient lift stops when ascending
CHANGE OF OWNERSHIP
The product can be reused. It must first be cleaned and disinfected in accordance with hygiene instructions. When
transferring the rollator, remember to hand over all the necessary technical documents to the new user. Beforehand, the
product must be inspected by an authorised specialist.
RECYCLING
When the product has become unusable and you have to throw it away: please contact your specialist dealer. If you wish
to carry out recycling for yourself, ask your local waste management company for disposal guidelines.
STORAGE
Failure to observe the storage conditions may lead to deterioration of the product and thus to the risk of
serious injury.
Do not store your product for an extended period of time near a heat source or in the sun (e.g. behind a window or near a
radiator) or near a cold source.
Keep away from all flames and sources of sparks.
Observe the conditions for storing and stowing your product.
Storing the lifting device in areas with very high humidity (baths, laundry, etc.) or in the open air, also the lack of hygiene
causes rapid loss of operating parameters and aesthetic of the product.
Press the emergency stop button for transportation or not using the patient life for a certain Of time. Tem-perature for
storage Of patient life shall be higher than O°C and relative moistures60%.
Inspection:
• In a dry and temperate place
• Protect your product by packaging it from dust, corrosion (e.g. abrasive elements, sand, sea water, salty air).
• Store all removed parts together in the same place (or locate them if necessary) to avoid mixing them with other
products when reassembling.
• All components must be stored without any load (do not place any objects that are too heavy on the product
elements, do not jam anything...).
DMI-F-04_Annexe 3_Notice 1.0 RAJOUTER CODE ERREUR
1. Check if the emergency stop button is pressed down.
2. Verify the controller and cables from actuator is properly con-
nected.
3. Check the battery level from display screen on the control
box.
4. Verify if there is any problem or damage on battery contact
panel.
5. Ask for after-sale service if patient lift still cannot work.
1. Check if the charger fully plugs into socket.
2. Verify if there is any problem or damage on battery contact
panel.
3. Ask for after-sale service if patient lift still cannot work.
1. Check if the emergency stop button is pressed down.
2. Employ the emergency descend device.
3. Ask for after-sale service if patient lift still cannot work.
20