TROUBLESHOOTING
MAIN
DISPLAY
Er 1
Er 2
Er 3
OvEr/Undr
SPECIFICATIONS
Display
pH
mV
Temperature °C
pH Temp Compensation
pH Buffer Recognition
pH Buffer Calibration Temp. Range
pH Electrode Offset Recognition
pH Electrode Slope Recognition
Power
Instrument Casing
Calibration Back-up
Audio Feedback
Autolock Feature
Operating Temp. Range
Operating Relative Humidity Range
ATC Probe
Dimensions (L x W x D)
Weight (batteries included)
POSSIBLE CAUSE
pH electrode offset is greater/less
than +/-1.5 pH.
STAND was pressed before the
electrode and ATC/Temp probe settle
to within ±1.5 pH of the buffer value.
pH electrode is faulty
pH electrode slope is off by more
than 30% of ideal slope.
SLOPE was pressed before the
electrode and ATC/Temp probe
settled to within 30% of the buffer
value.
Buffer 4.00, 4.01, 9.18 or 10.01 is not
correct.
Temperature is out of the 0.0 to 60.0
°C range.
Measured pH is out of the 16.00/-2.00
pH range.
Measured mV is out of the 1250/-
2000 mV range.
Measured temperature is out of the
-10/120 °C range.
Range
-2.00 to 16.00 pH
-1999 to 1999 mV
-10.0 to 120 °C
Accuracy
±0.1%, ±2 lsd
±0.1%, ±1 lsd
±0.5 °C
AUTO/MANual -10.0 to 120.0
(4.01, 7.00 & 10.01) or (4.00, 6.86 & 9.18)
°C
0 to 60
±90mV at pH 7.00 or 6.86
±30% at pH 4.00, 4.01, 9.18 or 10.01
One 9V battery/ Approximately 1000 hours
Waterproof IP 67
Yes
Yes, on all touch keys
Yes
°C
0 to 50
up to 95%
Thermistor, 10k / 25
18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 in x 1.5 in)
270 grams (.6 lb)
7
CORRECTIVE ACTION
Replace the buffer and/or
the pH electrode. Press ESC.
Allow sufficient time for the
electrode and ATC/Temp
probe to stabilize.
Return for service.
Check that the correct buffer
is used and that the
electrode slope is not off by
more than 30% from the
theoretical slope.
Allow sufficient time for the
electrode and ATC/Temp
probe to stabilize.
Replace the buffer and/or
the pH electrode. Press ESC.
Return for service.
Bring the buffer temperature
within range.
Return for service.
Bring the out of range unit
into the correct measuring
range.
If units are within proper
range, return product for
service.
Resolution
0.01 pH
1 mV
0.1 °C
°C
°C