7. Favourite key
The «favourite» key gives direct access to a predefined function, and can be
configured according to the needs of the user.
In order to assign a function to the «favourite» key, give a prolonged press on
and then select the required function :
Data transmission
0 dAtA
(default)
1 HoLd
Hold display
Re-initialise Min-
2 CLr
Max-Delta
Zero reset
3 SEt
Diameter calibration
(min.-max. mode)
Validation of selection: By a prolonged press on
Note :
a function can also be assigned via RS232 using the command <FCT
+ Function No.> / example : Change of Unit = <FCT4>
8. Adjustment of the measuring system
Certain applications need to adjust the instrument to the MIN (or MAX) measured
value. In this case, proceed as follows :
8.1 Adjustment of the instrument
•
Enter a Preset value corresponding to the actual size of the standard (see
chap. 3)
•
Select the MIN mode (or MAX, depending on the application) (see chap. 4)
•
Make a standard measurement (going through the turnaround point)
•
Adjust the instrument by selecting the SEt mode and pressing on the button
(see chap. 3)
•
The instrument is adjusted and ready to measure.
8.2 Measure
•
Make the measurements. The digital display stores and displays the MIN
(or MAX) measured value.
•
Before each new measurement, reset the measured value by a short press
on
SET
4 Unit
mm – IN
5 rEF
REF1 - REF2
6 PrE
Recall Preset
7 ModE
Nor–Min–Max-Delta
10
8 toL
10 dir
11 Mul
12 bt
, or a press on
Tol ON–Tol OFF
Measurement
sense, positive -
negative
Multiplier
Bluetooth
®
or
SET
MODE