Descargar Imprimir esta página

DeFelsko PosiTector GLS Manual De Instrucciones página 8

Ocultar thumbs Ver también para PosiTector GLS:

Publicidad

Ajustes Cal
Calibrate the PosiTector GLS probe to known standards by using included gloss
calibration standard (tile) or user supplied calibration standard. Place the probe on a
known standard and select the Calibrate menu item.
Alternatively, if Power-On Cal is selected in the Cal Settings menu the gage will
perform a calibration every time the gage is powered on.
Edit
Select a calibration standard from a list of saved standards to edit its Gloss Unit value
and name.
NOTA:
The calibration standard included within the protective base plate cannot be edited.
Abrir
Select a calibration standard from a list of saved standards. The calibration standard at
the top of the list is the one included within the protective base plate of the PosiTector
GLS probe.
New Cal Standard
Stores a new calibration standard into memory.
Adjust the Gloss Unit (GU) values using the on-screen keypad or by pressing (-) or (+).
Enter a name or serial number to identify the Standard using the on-screen QWERTY
touch keyboard.
Select OK, then follow the prompts to take a measurement of the desired gloss
calibration standard.
Borrar
Opción del Menú para configuración de Estadísticas y modos de alarma Hi/Lo.
Ver
Displays a list of stored standards and their Gloss Unit Values.
Power-On Cal
When selected, the Gage will perform a calibration every time the gage is powered-on,
using the calibration standard that was active when the probe was last used.
Page 6

Publicidad

loading