Possible Solution
Ensure insert is installed. Open the lid to the warmer and properly install
the insert. Press start. Refer to Insert Installation section on pages 11-12.
Close the lid and press start.
Check to see if plug in the back of the warmer is pushed completely in.
Unplug the device from the wall outlet and try another outlet.
Make sure that the IV pole adapter pin is engaged prior to installation
and snaps into place when the adapter is slid into the pole mount slot.
Allow milk to cool before feeding. Ensure proper settings are used. Note:
High temperatures may damage some active immunological properties
of breastmilk.
• Ensure that insert is installed properly. Refer to insert installation
section on pages 11-12.
• Use the proper size container in comparison to the volume of
breastmilk or supplement being warmed. Refer to warming and thawing
in a container section on pages 12-13.
• Place milk back into the unit within 60 seconds and press start to
continue warming. Do not initiate another warming cycle.
• Place milk back into unit within 60 seconds and press start to continue
warming. Do not initiate another warming cycle.
Ensure proper settings are used. Retrieve milk promptly when thawing
cycle is complete.
Ensure lid is closed and insert is fully seated.
Ensure insert is fully seated. If unit is near an intense infrared emitting
light source such as sunlight or heat lamps, move to another location.
Hold power button for at least 4 seconds or unplug the device, wait ten
seconds, plug the device back into the wall, and press the power button.
Contact Medela Customer Service.
Return for service.
Ensure fan is not blocked.
Turn off and unplug warmer, allow unit to cool down for up to 30
minutes. Warmer should resume operation. If Error Code 13 is
frequently generated, contact Medela Customer Service.
Ensure unit is in an area with adequate room temperature ventilation.
Ensure vents on bottom and rear of unit are not blocked.
21