Calibration Instruction
(This instruction is only for authorized personnel)
Please read the calibration instruction
very carefully.
Before calibrating you must have two 50 gram certified weights
and the extra foot that was provided or a small pin.
1
Scale must be on a level working area.
2
Remove the battery cover and detach the
aluminum plate near the batteries. The
calibration switch is underneath this plate.
3
Press "ON/TARE" button.
4
Wait for "0.0" to be displayed.
5
Using the extra foot or a pin, press and
hold the base calibration switch for 2
seconds. Release the base calibration
switch. The display will show "CAL".
Note: Do not put any stress on top platform when
you are calibrating the scale.
6
Immediately place the scale gently on a
level surface.
07
7
8
9
Aluminum plate
10
Calibration
switch
11
12
Press "ON/TARE" button again within 10 seconds after pressing the
calibration key. The scale will go into calibration mode.
Note: If "ON/TARE" button is not pressed within 10 seconds, the scale will turn off
automatically.
The LCD display will start showing some range of various numbers.
It will display "0000" with stabilize (°) mark on the left top corner.
Place 50g on the center of the platform within 15 seconds.
The display will show "0001" >>>"0004 and stabilize (°) mark, then
place an addition 50g within 15 seconds. (The total weight on the
platform is 100g)
The display will show "0005" >>> "0.000", then you can remove the
weights. The display will go to 0.0 and stabilize (°) mark.
Turn scale off. The calibration procedure is complete.
Note:
・When calibrating the scale, please pay close attention to the environment around you.
When doing the adjustments, please place the scale where there are no vibrations, wind,
temperature deviation, etc.
・"Err 4" appears on the display during calibration mode.
Please place weight(s) within 15 seconds.
Please make sure the scale is stable.
08