Never use abrasive cleaning agents, thinners or
benzene for cleaning and never immerse the
device in water or other cleaning liquids.
10. Guarantee
This device is covered by a 2 year guarantee from the date
of purchase. During this guarantee period, at our discre-
tion, Microlife will repair or replace the defective product
free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-
compliance with the instructions for use.
Damage caused by leaking batteries.
Damage caused by accident or misuse.
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Battery.
Should guarantee service be required, please contact the
dealer from where the product was purchased, or your local
Microlife service. You may contact your local Microlife
service through our website: www.microlife.com/support
Compensation is limited to the value of the product. The
guarantee will be granted if the complete product is
returned with the original invoice. Repair or replacement
Microlife OXY 200
within guarantee does not prolong or renew the guarantee
period. The legal claims and rights of consumers are not
limited by this guarantee.
11. Technical Specifications
Type:
Fingertip Pulse
Oximeter OXY 200
Display:
LED display
SpO
2
Measurement range:
70 ~ 100 %
Accuracy:
70 - 100 %: ±2 %
Resolution:
1 %
Pulse rate
Measurement range:
30 ~ 250 bpm
Accuracy:
30 ~ 99 bpm: ±2 bpm
Resolution:
1 bpm
Operating
5 - 40 °C / 41 - 104 °F
conditions:
15 - 80 % relative maximum
humidity
Storage conditions:
-10 - +50°C /
14 - 122 °F
10 - 93 % relative maximum
humidity
Automatic switch-off:
Automatically shut down in 8
seconds, when no or low signal
is detected.
Battery:
2 x 1.5 V alkaline batteries; size
AAA
Battery lifetime:
approx. 30 hours (using new
batteries)
Weight:
50 g (including batteries)
Dimensions:
58 x 32 x 34 mm
IP Class:
IP22
Reference to
EN ISO10993-1/-5/-10;
standards:
IEC 60601-1;
EN 60601-1-2;
ISO 80601-2-61;
EN 62304;
Expected service life:
5 years (when used 15 times/
day; 20 minutes for each
measurement)
Technical alterations reserved.
5
EN