Descargar Imprimir esta página

Braun Sensian Swipe 5 Manual Del Usuario página 9

Publicidad

Idiomas disponibles

Idiomas disponibles

Errors and troubleshooting
Error message
Situation
Error symbol: during taking
measurement
A "Err" sign appears when the
temperature measured is BELOW
93.2 °F or 34 ºC.
The mark "Err" appears if the
operating temperature is ABOVE 108
°F or 42.2 ºC.
When ambient temperature is too
high or too low to take an accurate
reading, device will provide an "Err"
message with flashing up/down
arrow icon. The error message will
appear after start up and memory
recall.
Low battery
When 20% of battery life is left, the
display FLASHES the low battery
warning symbol, however, the device
can still work until the battery life has
0% left.
Dead battery indication
If the steady battery icon is the
only symbol shown on the display,
the device cannot work.
Blank display
Thermometer does not have power.
All segments display or
Blank display
System error.
English
Solution
When you receive this symbol,
press the power button to reset the
thermometer before taking a new
measurement.
When you receive this symbol,
press the power button to reset the
thermometer before taking a new
measurement. If the ambient
temperature is not within the
allowed range of 59 ºF - 104.0 ºF
(15 ºC - 40 ºC), you will not be able
to take a reading.
Replace batteries. See instructions
on page 11.
Replace batteries. See instructions
on page 11.
Please check if the batteries have
been loaded correctly. Also check
polarity (<+> and <–>) of batteries.
Replace batteries. See instructions
on page 11.
If error persists, call consumer
relations.
7

Publicidad

loading

Este manual también es adecuado para:

Bst200us